DHG RoMedic ThoraxSling Instrucciones De Uso página 5

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43
Application of ThoraxSling on MiniLift
Application of ThoraxSling on MiniLift
Ask the patient to lean back against the vest before you start to raise the lift. Pay close attention to the patient.
When everything is adjusted and the patient is comfortable, raise the lift to the desired height. Discontinue raising if the
patient shows the least sign of discomfort.
ThoraxSling
Lifetime information & Periodic inspection
The product has an expected lifetime depending on how the product is used according to the manual instructions.
The service life of the product varies depending on usage frequency, washing, drying, etc.
Please be sure to read the product label and follow the care instructions for this product on www.directhealthcaregroup.com
The periodic inspection of the non-rigid body support unit should be performed at the timestamps stated by the
manufacturer, but at least every 6 months to be sure. More frequently inspections may be required if a non-rigid body
support unit is used or cleaned more frequently than advised by the manufacturer.
The periodic inspection instructions for this product are found on www.directhealthcaregroup.com.
I F U
Place the individually adjusted sit-to-stand lift in
front of the patient (see lift manual for instructions
for trial adjustment) and place the patient's feet on
the footrest. If a calf strap is to be used, it is ap-
plied first. Draw out the boom towards the patient
for easier
application of the strap loops.
Carefully place the lift as close as possible to the
patient, so as to determine which strap loop is the
most appropriate. Then, extend the straps before
hooking them to the slingbar.
ThoraxSling with seatsupport
5

Publicidad

Tabla de contenido
loading

Tabla de contenido