10. Calibration Procedure (cont.)
5. Press
shows "20 0--" and "g". Wait for
24
display to show "20 2--".
6. Place two 10 kg weights on the
platform after the display shows
"20 2--".
7. Wait for the display to show
"stabilized" mark (circle in the top
of left corner), then Press the
button to complete calibration.
8. Wait for display to show "func 1"
then remove the 20 kg weight.
9. Press
• If the following error messages are displayed, "Er0065"
or "Er0066", recalibration must be performed.
• Do not leave the calibration weights on the scale for
prolonged periods of time (more than a few minutes).
button again --> Display