PIPETTING RECOMMENDATIONS
Aspirate liquid into the pipette only when a
tip is attached to its tip cone.
Allow liquids, tips, and pipettes to equilibrate
to the ambient temperature.
Pre-rinsing of tip 5 times with the liquid to be
dispensed is recommended. This is important es-
pecially when dispensing liquids which have a
viscosity and density different from water.
Pre-rinse the tip several times before use when
pipetting liquids at temperature different from
ambient.
While pipetting, the pipette should be verti-
cally straight and tip should be dipped only a few
millimeters into the liquid.
Always control the push button movements
with the thumb for consistency.
Wipe the tip only if there is liquid on the out-
side of the tip, being careful to avoid touching the
tip's orifice.
Don't keep pipette in your hand while not
working, to avoid transferring of body heat
Use the correct pipette tip designed for use
with the particular pipette.
Select the correct pipetting technique (e.g. Re-
verse, forward etc.) depending on the nature of
the liquid.
Using excessive force to turn the push button
outside the range specified for it may jam the me-
chanism and damage the pipette.
MAINTENANCE
To maintain the best results from your pi-
pette, each unit should be checked every day
for cleanliness. Particular attention should be
paid to the tip cone(s).
This pipette has been designed for easy in-
house service. However, we also provide com-
plete repair service. Please return your pipette
to your local distributor for repair by Digipette
Technical Service. Before returning, please
make sure that it is free from any contamina-
tion.
Check the performance of your pipette re-
gularly e.g every 3 months and after every in-
house service or maintenance.
Cleaning Your Pipette
To clean your pipette, use ethanol and a soft
cloth or lint-free tissue. It is recommended to
clean the tip cone regularly..