c. Sodium bicarbonate solution (8- 10 %).
3. The following disinfection agents may be used:
a. Peracetic Acid Solution (0.08%).
b. Isopropyl Alcohol (70%).
c. Solution containing 70% Isopropyl Alcohol and 2%
chlorhexidine (e.g. Clinell Wipe).
d. Solution containing chlorine dioxide (e.g. Tristel Wipe).
e. PDI Sani-Cloth
Sani-Cloth
4. Blot dry the A-360 using an individual sterile surgical towel.
Caution:
• Do not autoclave.
• Do not rinse under running water.
• Do not soak in liquids.
• Avoid liquid or moisture from going into the A-360.
• Avoid touching the lens of the A-360.
• Be careful with A-360 touch screen.
22. OPERATING, STORAGE AND TRANSPORT
The A-360 should not be used, stored or transported at temperatures
o
below 0
C/32
not exceed 95 %. The air pressure must not exceed 500 to 1060 hPa.
Please handle with care as with any other electronic device. Use the
storage case provided to protect the device between each use.
23. TECHNICAL DATA
24.1. A-360
Electrical supply input ratings: 5 V DC, 1.5 A
Supply Voltage: Rechargeable Li-Po 3.7 V, 2100mAh.
Baseband Modulation – WiFi - 802.11b: DSSS
Output RF Power: +17.5 dBm Typ.
Transmission range: 10 m (approx).
Screen : 2.8" TFT-LCD, 240 x 320 RGB.
Touch Panel : OGS type 2.8" Projected Capacitive Touch Panel.
Video Sensor: Digital CMOS video sensor, VGA 640 x 480.
Video resolution: 432 x 480.
Dimensions: 58 mm x 82 mm x 93 mm
Weight: 220 grams.
24.2. Stand-alone Battery Charger
Electrical supply input ratings: 100-240 V ~, 50-60 Hz, 0.1-0.2 A