Capacitance measurement / F
Set the selector switch to "
lead to the "COM" jack and the red test lead to the "V/Ω" jack. Connect the test leads with the
test object and read off the measurement result from the display.
Measurement range
60 nF
600 nF
6 µF
60 µF
600 µF
6 mF
60 mF
Discharge the capacitors before every measurement!
Automatic switch-off
The instrument switches off after c. 15 minutes (APO). Continuous actuation
of the "APO" button switches the Auto Power Off function on or off.
Backlighting
Continuous pressing of the "HOLD" button switches on the backlighting.
The illumination switches off automatically after c. 1 minute.
Saving function (HOLD)
At difficult measurement positions, the display cannot perhaps be read correctly.
Pressing the "HOLD" button freezes the current measured value which can then
be read in comfort.
Repeated pressing of the "HOLD" button returns to normal measuring mode.
True RMS
If measuring non-sinus wave forms, low measurement errors can occur if using the True RMS
function, in comparison to using traditional measurement procedures.
Sinusform and non-sinus form signals can be accurately measured using the True RMS function.
24
2
Operation
/
". Actuate the "SEL" button x 3. Connect the black test
Resolution
10 pF
100 pF
1 nF
10 nF
100 nF
1 µF
10 µF
Electrical energy may cause serious injuries!
Always wear personal protective equipment!
^ hazet.com • ] info@hazet.de
Accuracy
± 4 % + 3 digit