HCB-200-650F Specification
The figures stated in the specification are as measured 0-50 mm from the base of the well of the insert,
using an SPRT, at the time of calibration.
Minimum temperature
Maximum temperature
Temperature accuracy
Temperature uniformity
Temperature stability after 10 minutes
Display resolution
Set point display resolution
Heat up 35 minutes
Cool down 30 minutes
Fan cooling Automatic
Dimensions Height x Width x Depth
Large well in the block:
Weight
Options
Central Well for inserts, see the list of inserts
Full bi-directional RS232 Comms Port 9way D type plus software and cable
Soft carry case
Hard carry case
UKAS calibration certificate
25°C above ambient
650°C
±0.4°C
±1°C
±0.09°C
0.01°C
0.1°C
20°C to 600°C
600°C to 200°C
Automatic
278mm x 170mm x 300mm
Depth / Diameter 152mm / 38.2mm
9.6Kg
16