6.3 Biomimetic Adjustment
NB: Carry out static alignment while ensuring the user has some means of support such as
parallel bars. This is standing alignment only.
No
Does the user report
a feeling of falling
forwards?
Yes
Shift the foot forwards
slightly relative to the
socket (e.g. tilt using
proximal & distal
interfaces)
Use shift for static alignment and standing.
The device should encourage some degree of self adjustment to achieve a sense of balance for
the user during standing.
All manuals and user guides at all-guides.com
Check bench alignment taking
heel height into account
Ensure PF and DF valves are set as shown.
(the user must feel some movement in
the ankle)
Ask the user to stand with weight
evenly distributed between each foot
Does the user feel sufficiently stable
using minimal muscle control?
No
Does the user report a
feeling of falling backwards?
Shift the foot backwards slightly
relative to the socket (e.g. tilt
using proximal & distal interfaces)
}
Yes
Can the user dorsiflex the
ankle by approx. 3°?
Yes
accustomed to the ankle for
Yes
10
No
Adjust the angle
of the foot at the
distal pyramid
interface
Allow user to become
10 mins before carrying
out any DF/PF valve
adjustments
Proceed to
Dynamic
Adjustment
938367/1-0816