Instructions for Use
Bipolar Sealing System marSeal 5 with sterile blade holder
NOTICE
• Dried or fixated operation residues may corrode the surface of the products.
— Ensure a short disposal time.
— Do not used fixating pre-cleaning, nor fixating disinfecting agents.
• Unsuitable cleaning agents or excessive temperatures may result in damage to the
product.
— Use only cleaning and disinfecting agents that have been approved for use with
plastics and stainless steel.
— Use only approved cleaning and disinfecting agents according to the manufacturer's
instructions
— Observe the information regarding concentration, temperature and exposure time.
• Use of hydrogen peroxide can lead to premature wear of the instrument and thus reduce
its lifetime.
• Additional reprocessing information is available at http://kls-martin.com/phnet/
Revision 2
63