Descargar Imprimir esta página

Publicidad

Idiomas disponibles

Idiomas disponibles

Enlaces rápidos

EM720
Hi-Res Differential Manometer
INSTRUCTION MANUAL
ENGLISH
ISED
1-800-547-5740
www.ueitest.com • email: info@ueitest.com
WARRANTY
The EM720 is warranted to be free from defects in materials
and workmanship for a period of one year from the date
of purchase. If within the warranty period your instrument
should become inoperative from such defects, the unit will
be repaired or replaced at UEi's option. This warranty covers
normal use and does not cover damage which occurs in
shipment or failure which results from alteration, tampering,
accident, misuse, abuse, neglect or improper maintenance.
Batteries and consequential damage resulting from failed
batteries are not covered by warranty.
Any implied warranties, including but not limited to implied
warranties of merchantability and fitness for a particular
purpose, are limited to the express warranty. UEi shall not
be liable for loss of use of the instrument or other incidental
or consequential damages, expenses, or economic loss,
or for any claim or claims for such damage, expenses or
economic loss.
A purchase receipt or other proof of original purchase date
will be required before warranty repairs will be rendered.
Instruments out of warranty will be repaired (when
repairable) for a service charge.
For more information on warranty and service, contact:
www.ueitest.com • Email: info@ueitest.com
1-800-547-5740
This warranty gives you specific legal rights. You may also
have other rights, which vary from state to state.
FEATURES
• Measurement range: -80 to +80 inH
2
• Differential Pressure
• Units of Measure: bar, ozin
, psi, inHg, mbar, mmHg,
2
Pa(kPa), kgcm
, inH
O, ftH
O, cmH
2
2
2
2
• Magnetic Mount
• Backlight
• HOLD
• MIN/MAX/AVG Capture
• Ambient temperature measurement
• Auto Power off
• Low Battery Indicator
• Zero Button
Copyright © 2022 Kane USA Inc. All rights reserved.
WARNING
Read entire Safety Notes section regarding potential hazard
and proper instructions before using this meter. In this manual
the word "WARNING" is used to indicate conditions or
actions that may pose physical hazards to the user. The word
"CAUTION" is used to indicate conditions or actions that may
damage this instrument.
• To avoid false readings, replace the batteries if a low
battery indicator appears.
WARNING
• Do not exceed Maximum Over Pressure of ±10 PSI.
WARNING
• Do not use the manometer if it operates abnormally.
• Always test the manometer before use to ensure it is
operating properly.
• Always ensure that the connection hoses you are using are
free from kinks, blockage or splits.
• Always double-check all connections before testing.
• While using the manometer, if you suspect gas is leaking,
or if you smell gas, close off gas supply and ventilate the
occupied space. Identify and repair the source of the
leak before continuing with the test.
• Never connect the manometer to an unknown pressure
source if the source is twice the working pressure of the
instrument; the sensor may be damaged and it may
cause injury to the user.
• Never open the manometer or attempt to modify it in any
way. Otherwise, accident and injury may occur. And the
warranty is voided.
• These manometers do not measure water pressure; under
these conditions they will fail.
WARNING
This manometer is designed for trained trade professionals
who are familiar with the hazards of their trade. Observe
all recommended safety procedures and use of personal
protective equipment.
MIN AVG MAX
Operating Temperature: 32°F to 104°F/ 0°C to 40°C
Storage Temperature: 14°F to 122°F/ -10°C to 50°C
Operating Humidity: maximum relative humidity 80%
for temperatures up to 88°F (31°C) decreasing linearly to
50% relative humidity at 104°F (40°C).
Operating Altitude: Up to 2000m
Pollution Degree: 2
Overvoltage Category: I
Over-range: OL is displayed
Apo: Auto Power off after 30 minutes
Dimensions: 2.95 x 6.25 x 1.41 in
0
Item Weight: 0.65 lbs
Calibration: Recommended Annually
O
Certifications: UKCA, CE, FCC, RoHS, REACH Compliant,
ISED (CAN ICES-003(B) / NMB-003(B))
Rated Power: 4.5 VDC (3X1.5 VDC, LR6, AA type batteries)
Accuracy: ±0.03 < 3 inH
Resolution: 0.001"(-9.999 to +9.999 inH
0.01" (< -10, > +10 inH
IMPORTANT SAFETY WARNINGS
SYMBOLS
Hold/Capture Value
Maximum measured value displayed
Minimum measured value displayed
Timed mean average calculated
values displayed
Record measured values
Differential Pressure
Warning or Caution
Negative Pressure
Battery Indicator
GENERAL SPECIFICATIONS
O, ±1%rdg > 3 inH
O
2
2
O),
2
O)
2
OVERVIEW AND OPERATING INSTRUCTIONS
A
H
B
I
C
D
F
E
J
G
A. Pressure Sensor Inputs
• Left Connection: Negative pressure input
• Right Connection: Positive pressure input
B. Display with Backlight
1. Hold: Values are captured/held
2. Dif: Differential Pressure
3. REC: The manometer will record the Minimum
pressure, the Maximum pressure, and
Average pressure
4. MAX/MIN/AVG
- MAX: Maximum value captured is displayed
- MIN: Minimum value captured is displayed
- AVG: Timed mean value calculation is displayed
NOTE: REC/MAX/MIN/AVG function does not work
when in Differential Pressure test mode
5. Battery: When battery is too low for safe operation,
the Low Battery indicator will display
See Battery Replacement section at the end of
this manual.
6. Time: (H/M/S) Hours/Minutes/Seconds
C. DIF ZERO Button
• Press to start Differential Pressure test. Press to
stop Differential pressure test.
• To zero the Manometer, press and hold for 4 seconds.
D. POWER Button (Red)
• Press to power on manometer. Press to power
off manometer.
E. HOLD BACKLIGHT Button
• Press to hold pressure reading on display.
• Press and hold to turn on display backlight. Back
light has a 60 second duration.
F. REC MAX/MIN Button
• Press to record MIN/MAX/AVG values
• Press and hold to exit.
G. UNITS Button
• 11 Selectable units of measurement
H. Magnetic Mount
I. Battery Compartment Cover
J. Serial Number
SPECIFICATIONS
Units
Range (Max.)
Resolution
bar
±0.199
0.001
ozin
±46.24
0.01
2
psi
±2.890
0.001
InHg
±5.884
0.001
mbar
±199.3
0.001
mmHg
±149.5
0.1
Pa/kPa
±19.93 kPa
0.1 Pa
kgcm
±0.203
0.001
2
inH
O
±80.00
0.001
2
ftH
O
±6.667
0.001
2
cmH
O
±203.2
0.1
2
.
Accuracy
± 0.03 < 3 inH
O,
2
±1% rdg > 3 inH
O)
2
202204 • 0322

Publicidad

loading

Resumen de contenidos para UEi Test Instruments EM720

  • Página 1 See Battery Replacement section at the end of protective equipment. this manual. The EM720 is warranted to be free from defects in materials 6. Time: (H/M/S) Hours/Minutes/Seconds and workmanship for a period of one year from the date SYMBOLS of purchase.
  • Página 2 In. of H 2.538217706 Caution the user that changes or modifications not expressly approved by the party responsible for I. EM720 TEST SOURCES compliance could void the user’s authority to operate the equipment. A. Positive Pressure Input B. Negative Pressure Input CLEANING C.
  • Página 3 Low capacitados que están familiarizados con los riesgos Battery (batería baja) en pantalla El manómetro EM720 está garantizado contra defectos en del oficio. Siga todos los procedimientos de seguridad Consulte la sección Reemplazo de la batería al final materiales y mano de obra durante un período de un año a partir de...
  • Página 4 Adviértale al usuario que los cambios o modificaciones que no 2.538217706 sean aprobados expresamente por la parte responsable del cumplimiento, podrían anular la autorización al usuario para I. FUENTES DE PRUEBA DEL MANÓMETRO EM720 operar el equipo. A. Entrada de presión positiva LIMPIEZA B.