Descargar Imprimir esta página

Steinberg Systems SBS-LZ-300D Manual De Instrucciones página 26

Ocultar thumbs Ver también para SBS-LZ-300D:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 80
5. Set the unit parameters:
rotation speed
Press MENU button (B), the speed value indication will start flashing, use
buttons (C) and (D) to set the speed value.
Caution: The value indicated on the display should be multiplied x10 to
obtain the real speed value
spin time
Press Menu button once or twice depending on the preferred time unit -
seconds or minutes. Using buttons (C) and (D) set the speed value.
6. The maximum centrifugal force (RCF) value can be determined from the
formula below:
RCF= 11.18* (n/1000) *2r
Where:
r - radius of the centrifuge expressed in cm
n- centrifuge rotational speed in rpm
7. Press Start/Stop (E) button to start the unit. Caution: During the first start a
slight friction sound can be heard for the first 2-3 minutes.
8. If error E3 appears on the machine display check if the cover is properly
closed.
9. To stop the machine press the Start/Stop button (E) or wait until the timer
counts down to 0.
The unit will not come to a complete stop until the current speed is 0 in the
display.
10. Remove the tubes with the centrifuged samples.
11. Turn off the unit by pressing the I/O button. Disconnect the unit from the
power source when not in use.
3.4.
Cleaning and maintenance
a)
Pull out the mains plug before each cleaning and when the unit is not in use.
b)
Use only non-corrosive cleaning agents for cleaning the surfaces.
c)
After each cleaning, all the parts should be dried well before the unit is used
again.
d)
Store the unit in a dry and cool place protected from moisture and direct
sunlight.
e)
Do not spray the unit with a stream of water or immerse it in water.

Publicidad

loading

Este manual también es adecuado para:

Ex1003072