Quick Reference Guide – New MA 41
Pure-Tone Audiometry
1.
Select TONE on right side of screen
2.
Press L or R to select test ear using key on
bottom left of screen.
3.
Select transducer
4.
Select frequencies using +/- key
5.
Change intensity with the SIDE DIALS
Push the BLUE BUTTON next to the side
6.
dials to present the stimulus.
7.
Press S key to save the threshold
Masking
1.
Turn masking noise on or off using button
at bottom right of screen
2.
STIM light will remain on to show that
masking noise is present
3.
Adjust noise level using SIDE DIAL
4.
Press S key to store threshold.
NOTE: Masked symbol will be plotted when masking
noise is on and above 0 dB HL.
Speech Audiometry
1.
Select SPEECH on right side of screen
2.
Press L or R to select test ear using key on
bottom left of screen.
3.
Select type of speech test: SRT, WRS, or
UCL.
4.
Select input: MIC, CD, or WAVE
5.
Change level using SIDE DIALS
6.
Score words using +/- key
7.
Press S key to save result
1162-1208 Rev B
MAICO Diagnostics
All manuals and user guides at all-guides.com
Calibrating VU for Mic and CD Input
1.
Select Mic or CD
2.
Press MONITOR at left of screen
3.
Press INCAL at right of screen
4.
Play input: voice or cal tone on CD
5.
Use SIDE DIAL to adjust input signal until VU
is green.
6.
Press OK
7.
Press OK again
Using Wavefiles
1.
Select SPEECH
2.
Select type of speech test: SRT, WRS, UCL,
MHA
3.
Select WAVE
4.
To expand a menu category (Adult, child,
etc.), press the BLUE BUTTON
5.
Use the SIDE DIAL to scroll down the menu
and highlight the desired list
6.
Press the BLUE BUTTON to select
7.
Change presentation level using SIDE DIALS
8.
Press PLAY to start the list
9.
Use + /– keys to score word responses
10. Press S key to store result
Talk Forward
1.
Press and hold STIM TALK button
Adjust level of your voice using SIDE DIALS
2.
while STIM TALK button is pressed
7625 Golden Triangle Drive, Eden Prairie, MN 55344 Tel: 888.941.4201
Monitors & Talk Back
1.
Press MONITOR on the left side of screen
2.
Adjust monitor of Channel 1 and 2 using the
SIDE DIALS
3.
Adjust the talkback using the +/- key
4.
Preferences for monitors and talkback can
be changed using the soft keys at the
bottom of the screen.
NOTE: The MONITOR key is not accessible while in
the WAVE screen. It is recommended to set up
monitors and talkback levels at the beginning of the
test session.
Saving/Printing Test Results
1.
Press MENU key
2.
Press PDF to save results to a PDF file on
the USB stick.
OR
3.
Press PRINT to print the results to an
approved connected printer
OR
4.
Press PATIENT to save the results under a
patient name in the unit.
NOTE: If results are saved under PATIENT, results
will still need to be saved as a PDF in order to print
out a paper copy. The patient's name will then be
included on the printout.
www.maico-diagnostics.com
pg 1/2 12/12