Descargar Imprimir esta página

KLS Martin marVac Instrucciones De Uso página 40

Detector de corriente de red externa idetect

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Make sure that the maximum power consumption of the connected electrosurgery or laser surgery device
does not exceed 10 A! The smoke evacuator and the connected electrosurgery or laser surgery device
must be operated via 2 separate mains supplies.
40
marVac® / maXium® smart Vac iDetect external line current detector
• The line current detector is attached to the back of the
®
marVac
/maXium smart Vac using the four Phillips
screws (1) provided.
This requires the plastic plugs in the thread grooves to
be removed first.
• Connect the MCB connection cable (2) to the 9-pin Sub-
D socket (marked with MCB) of the smoke evacuator.
This cable supplies the power for the line current
detector
and
communications
evacuator.
• Plug the non-heating device cable (3) located on the line
current detector into the non-heating device socket of
the electrosurgery or laser surgery device.
Instructions for Use
with
the
smoke
Revision 04

Publicidad

loading

Productos relacionados para KLS Martin marVac

Este manual también es adecuado para:

MaxiumSmart vac