ISO 80601-2-70:2015 Medical electrical equipment - Part 2-70: Particular
requirements for basic safety and essential performance of Sleep apnea
breathing therapy equipment.
ISO 8185:2009 Respiratory tract humidifiers for medical use - Particular
requirements for respiratory humidification systems.
IEC 60601-1-11:2015 Medical electrical equipment -- Part 1-11: General
requirements for basic safety and essential performance –Collateral Standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment.
Électrique
Alimentation spécifiée: Resvent Medical Technology Co, Ltd.
Modèle: MDA60-220S24-W1
Entrée ca: 100-240V ~ 1.18-0.56A 50/60Hz 2A Max
Sortie DC: 24 V⎓2.5 A
Alimentation spécifiée: SHANGHAI BIAOJUN ELECTRONICS
TECHNOLOGY Co., Ltd.
Modèle: BJE01-40-006HM
Entrée ca: 100-240V ~ 50/60Hz 2A Max
Sortie DC: 24V⎓2.5 A
Spécifications de sécurité
Class II
Type BF
Ingress Protection: IP22
Filtre à air
Filtre à air: Efficacité du filtre: >75% (7 micron dust)
Pression
Plage de réglage: 4-20 cmH
Max Pression constante à un seul défaut: 40 cmH
Précision de contrôle de pression: ±0.5 cmH
Variation de pression dynamique: < 1 cmH
Contrôle de pression Précision: ± (2% * lecture pleine échelle + 4% * lecture
réelle)
O
2
O (Incertitude élargie: 1,87%)
2
O (Incertitude élargie: 2.55%)
2
33
O
2