TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on
Batteries have
when the START/STOP
run down.
button is pushed.
Battery polarities have
been positioned incor-
rectly.
EE measurement error
Cuff has been placed
symbol shown on display
incorrectly.
or the blood pressure
value is displayed exces-
sively low
Did you talk or
(or high).
move during
measurement?
Shaking of the wrist with
the cuff on.
E1 error symbol shown
Air circuit
on display
abnormality.
E2 error symbol shown
Inflation pressure exceeding
on display
300 mmHg.
E3 error symbol shown
Error determining mea-
on display
surement data.
EP error symbol shown
System error.
on display
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance
should you disassemble or attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on the warranty page.
CORRECTION
Replace them
with two new "AAA" alkaline
batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff properly so that it
is positioned correctly. Measure
again.
Keep wrist steady during
measurement. Measure again.
Measure again.
Switch the unit off, then mea-
sure again.
Measure again.
Measure again.
31
SPECIFICATIONS
Power Source:
Two 1.5V DC "AAA" Alkaline batteries
Measurement
Oscillometric
Method:
Measurement
Pressure: 40~280 millimeters Mercury (mmHg)
Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity: 60 memories for each User (120 total)
Auto-shut-off:
1 minute after last button operation
Operation
Temperature: 41°F~104°F (5°C~40°C)
Environment:
Humidity: 15% ~ 93% RH
Storage/
Temperature: -13°F~158°F (-25°C~70°C)
Transportation
Humidity: Less than 93% RH
Environment:
32