5. Wet disks should be dried thoroughly with a lint free wipe. Handle the
disk by the edges. Between tests store the disk on a flat surface. Do
not agitate a partially filled disk to avoid leaking and contamination.
Partially used disks should be discarded at the end of the day.
6. Disks should be filled and used within 10 minutes. They cannot be filled ahead of time.
For filling and troubleshooting tips go to:
GENERAL OPERATING PROCEDURES
METER When a filled disk is placed in the chamber and the lid is closed, the meter spins at high
speed to distribute the sample to the test wells. Next the meter slows to maximize the pumping
action of the stainless steel mixing beads as the reagents mix with the sample water. Each
reaction is then read at the proper time and wavelength for that reagent system.
The button located in the lower center of the top of the meter turns the instrument on
and off.
The Blue indicator light of the on/off button indicates the status of the instrument.
Steady blue light – the blue LED will remain steady to indicate that the meter is on and ready to
run a test.
Blinking blue light (three blinks/second) – a test is in progress and the disk is spinning. Do not
open the lid when the disk is spinning.
Care should be taken when closing the lid. Do not slam the lid. Wiring between the lid and the
body of the photometer passes through the hinge. The meter will not run with the lid open.
TOUCHSCREEN DISPLAY
When the button is pressed to turn the meter on,
the Test Screen will be displayed.
The display screen is touch-activated. To make a
selection, tap the icon or word on the screen with
a fingertip, fingernail, pencil eraser, or stylus.
¬ Gently wipe smudges from the screen with the
Cloth Wipe (3580-WIPE).
¬ Do not touch the screen with a sharp object.
¬ Do not place objects on the screen that will
scratch or damage it.
¬ Avoid touching the screen with wet fingers.
TESTING
1. Press
and hold until the meter turns on.
2. Tap
. Select a sanitizer type. Tap
confirm.
3. Tap
. Select a disk series (found on disk packaging). Tap
NOTE: Disk Series are limited by Sanitizer selection.
4. Remove a disk from the packaging.
5. Use the syringe (1189) to fill the disk with the pool or spa water sample.
6. Insert the disk. Cover the disk with the Universal Disk Cover (1719). Close the lid.
7. Tap
to start test. Tap
8. The results will be displayed.
9. Choose an option.
¬ Tap
to add a tag.
¬ Tap the highlighted
¬ Tap the hightlighted
¬ Tap the highlighted
¬ Tap
to return to the Test Screen.
10. Press and hold
lamotte.com/spin_support
to cancel the test. If the test is cancelled discard the disk.
to save the test results to the test log if Auto Save is not enabled.
to send the results to the enabled Mobile Bluetooth Printer.
to send the results to a Bluetooth enabled device.
for 2 seconds to turn the meter off.
Sanitizer
Disk
Type
Power/
Battery
Status
to
5
Start
Test
Sanitizer
Salt
Tap to
Start Test
Disk Series
203
16-Apr-20
12:14 PM
Date &
Test
Time
History
to confirm.
Settings