4.
BATTERY OPERATION
FUEL GAUGE: Each LLS-2054BE battery is equipped with a fuel gauge that displays the relative
charge state of the battery. To activate the fuel gauge, press the small circular button next to the
gauge. Each LED on the fuel gauge represents 25% of the fully charged capacity. Below is the fuel
gauge legend:
•
All four LEDs illuminated – Capacity = 76 – 100%
•
Three LEDs illuminated – Capacity = 51 – 75%
•
Two LEDs illuminated – Capacity = 26 – 50%
•
One LED illuminated – Capacity = 10 – 25%
•
One blinking LED – Capacity < 10%
CHARGING: The LLS-2054BE should only be charged with ST Technologies
battery chargers. To charge, orient the battery to line up the charger and battery contacts, and
place the battery into the slot. Do not force the battery into the slot. Refer to the charger IFU for a
detailed operation.
ABOUT SMBus and SBS: The Smart Battery System defines the parameters that are stored by
an SBS-compliant battery. These parameters include full battery status and fuel gauging
information. The System Management Bus is the language by which these parameters are
communicated between the battery, the charger and the host device.
NOTE: The LLS-2054BE no longer requires fuel gauge re-calibration.
5.
CLEANING INSTRUCTIONS
•
Remove the battery from the host device.
•
Clean the exterior of the battery with a cloth that has been dampened (not dripping) using a
hospital disinfectant. Follow applicable hospital protocols during cleaning and dry immediately.
6.
USER SERVICE
The LLS-2054BE has no user or field serviceable components. It can only be serviced at the
factory by factory trained technicians.
7.
REPLACEMENT PARTS
For replacement parts, additional or optional equipment :
Please call Customer Service @ 877-814-2237
®
LIT-234 CUDA
Surgical
Rev. D
Figure 1. The LLS-2054BE Battery
Page 5 of 36
®
supplied smart
(English)