Klarstein LND018915V3 Manual De Instrucciones página 19

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Guidance and manufacturer's declaration – electromagnetic emissions
The „SAD light" is intended for use in the electromagnetic environment specified below. The customer or the user
of the „SAD light" should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
portable and mobile RF communications equipment and the Medical „SAD light"
The „SAD light" is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Medical „SAD light" can help prevent electromagnetic interfe-
rence
by maintaining a minimum distance between portable and mobile RF communications equipment (trans-
mitters) and the appliance as recommended below, according to the maximum output power of the com-
munications equipment.
Rated maximum
output power
of transmitter W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
19
Compliance
Group 1
Class B
Class A
Complies
Recommended separation distances between
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
0,12
0,38
1,2
3,8
12
Electromagnetic environment – guidance
The „SAD light" uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The „SAD light" is suitable for use in all es-
tablishments, including domestic establish-
ments and those directly connected to the
public low-voltage power supply network
that supplies buildings used for domestic
purposes.
80 MHz to 800 MHz
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
0,23
0,73
2,3
7,3
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

10030744

Tabla de contenido