At the end of the third calibration point, the instrument returns automatically to measure mode.
To perform a one- or two-point calibration, press the key
point.
Note: electrode calibration is an essential operation for the quality and truthfulness of a
measurement. Therefore, make sure that the buffers used are new, unpolluted and
at the same temperature
ATTENTION: Before proceeding with the calibration operations, carefully consult the safety data
sheets of the substances involved:
Calibration buffer solutions.
STORAGE solution for pH electrodes.
Filling solution for pH electrodes.

• Calibration with manual values

Example
two- point calibration pH 6.79 e pH 4.65 (DIN19267).
:
Access the Setup menu for pH and select in P1.1 Custom, press twice the key
measurement and position in pH mode
Press
to enter the calibration mode.
Rinse the electrode with distilled water and gently dab it with paper towel. Dip the electrode in the
first pH buffer solution (e.g. pH 6.79).
Wait for the pH value on the display to stabilize; when the icon
modify it using the directional keys by entering the correct one (e.g. pH 6.79), as suggested by the
string "ADJUST THE VALUE" and by the icon
Note: Check the buffer value according to the temperature.
When the icon
measured value flashes on the display and the beaker icon appears with the buffer identification
colour
.
Remove the electrode, rinse with distilled water, dab it gently with paper towel and dip it in the next
buffer (e.g. pH 4.65).
Wait for the pH value on the display to stabilize; when the icon
flashes, modify it using the directional keys by entering the correct one (e.g. pH 4.65), as suggested
by the string "ADJUST THE VALUE" and by the icon
When the icon
measured value flashes on the display, the Slope% and, next to the first beaker, the icon with the
identifying colour of the second buffer
At the end of the second calibration point, the instrument automatically returns to measure mode.
To perform a one-point calibration just press the key
Note: If you are working with manual temperature compensation (MTC), update the value before
calibrating the instrument.
Series 7 Vio ENG Version 2.0 April 2020
.
.
appears again, press the key
appears again, press the key
appears.
20
once finished the first or second
appears and the value flashes,
to confirm the first point; the actual
appears and the actual value
.
to confirm the second point; the actual
after finishing the first point.