8. Specifications
Measuring ranges ≥ 6 ... < 40 bar:
Measuring ranges
EN
0 ... 6 bar
0 ... 10 bar
0 ... 16 bar
0 ... 25 bar
14
Temperature limit
20 °C [68 °F]
≤ ±1,7 %
≤ ±1,3 %
≤ ±1,0 %
≤ ±1,0 %
50 °C [122 °F]
≤ ±3,4 %
≤ ±2,3 %
≤ ±1,6 %
≤ ±1,3 %
WIKA operating instructions pressure sensor, model MH-4
80 °C [176 °F]
≤ ±5,8 %
≤ ±3,8 %
≤ ±2,7 %
≤ ±2,1 %