Maintenance of the byko-spectra
Cleaning
When to Relamp the byko-spectra
Relamping Procedure
Maintenance of the byko-spectra
NOTE: Disconnect the power cord before
cleaning or relamping the byko-spectra!
The viewing area of the byko-spectra should be
kept clean and free of samples, notices, etc. To
clean the neutral gray surface, use a soft, damp
cloth with a mild soap or detergent. For stubborn
marks or stains, use a non-abrasive "soft-scrub"
type of cleaner. DO NOT use organic solvents on
the surface.
In order to maintain proper color quality and light
intensity, the lamps should be replaced when the
time expires for each individual lamp. Each lamp's
use is tracked by the Colorguard which does all
the timing and warns you of the lamps need for
replacement. D65, CWF, and TL84 sources are to
be replaced at 2500 hours, Inc is to be replaced
at 1750 hours and U.V. is to be replaced at 500
hours. To order replacement lamps for your
viewing system, refer to the product label on the
rear of the unit (located near power cord) for a
relamp kit code.
Rotate the lamps one-quarter turn and remove
them from their lampholders. Clean the reflector
surfaces and install the new lamps, rotating them
one-quarter turn to lock them in place. After
relamping, and when the unit is turned on, select
the "relamped" source and depress and hold the
reset switch located beneath the hole next to the
switch board (as shown by arrow) until the unit
beeps three (3) times. This is best accomplished
with the use of a straightened paper clip. The timer
is now reset back to "0". Repeat operation for
each "relamped" source.