Temperature-taking hints
ALWAYS replace disposable hygienic lens filters to maintain accuracy and hygiene . A measurement taken in
the right ear may differ from a measurement taken in the left ear . Therefore, ALWAYS take the temperature
in the same ear . The ear must be free from obstructions or excess earwax build-up to take an accurate
reading .
External factors may influence ear temperatures, including when an individual has:
Factor
Poor probe placement
Used hygienic lens filter
Dirty lens
In the cases below, wait 20 minutes prior to taking a temperature .
Extreme hot and cold room temperature
Hearing aid
Use the untreated ear if ear drops or other ear medications have been placed in the ear canal .
If an individual has been laying down on their side with an ear face down on a pillow, use the
ear that has been facing upwards .
Last reading recall
1 . The last temperature taken is stored in the memory and
will be automatically displayed for 5 seconds when it is
turned on again .
How to illuminate the display
The thermometer includes a convenient night light to
illuminate the display in a dim environment .
The light will turn on when you press any button . It will stay
on until the thermometer is inactive for up to 15 seconds
even after temperature is taken .