CLEANING AND MAINTENANCE
The device requires no particular maintenance, it should simply be
cleaned regularly.
Remove the plug before carrying out any cleaning operations.
NEVER IMMERSE THE APPLIANCE IN WATER.
When the application is finished, the ampoule and the accessories have
to be sterilized with denatured alcohol or with a solution based on
hypochlorite which is easy available in pharmacy.
The outside of the appliance and the storage space should be cleaned
when necessary with a soft cloth soaked in alcohol. Before using the
appliance, make sure the alcohol has completely evaporated.
Do not use other substances, liquids or excessively damp cloths,
as any seepage into the appliance could cause irreversible damage.
Problem
1. Poor nebulization
2. Poor nebulization
3. Lack of nebulization
4. Lack of nebulization
5. Slow nebulization
6. Noisy apparatus
Cause
Obstructed cruet
Obstructed cruet
Poorly fitted nozzle
Obstructed cruet
Very oily medication
Prolonged use
Solution
Proceed with cleaning and disinfecting the
cruet as indicated in the manual
If the cleaning did not resolve the problem,
replace the cruet
Using a finger, firmly press the nozzle
(cylindrical sleeve) located inside the
polycarbonate cruet base
Replace the cruet
Dilute the medication with physiological solution
Apply to the reseller or to the
ARDES service center