Cleaning
Cleaning of this instrument should be carried out using a soft cleaning anti-static
liquid, after verifying that it doesn't affect the plastic parts used in the case and
in the Control Panel of this equipment.
Replacement fuse
To check the instrument FUSE , remove it with a screw driver. If the fuse is
ruptured replace it by another with the following specifications:
Fuse Schurter, model SPT 5x20 (Time-lag) 5A/250V. High breaking capacity H.
Technical specifications
Test currents
Resistance ranges
Resolution
Output voltage
Measurement principle
Basic accuracy
Advanced features
Serial data output
Environmental protection
:
1 mA - 10 mA - 100 mA - 1 A - 10 A. Each current may
be continuously adjustable from 0 to 100%.
:
0-2000 µΩ @ 10 A
0-20 mΩ @ 10 A
0-200 mΩ @ 1 A
0-2000 mΩ @ 100 mA
0-20 Ω @ 10 mA
0-200 Ω @ 1 mA
:
1 µΩ @ 10 A
:
Up to 10 Vc.c. @ 1 A (open circuit)
:
Four-terminal, Kelvin-type.
:
±0,2 % of reading ± 2 digits
:
Digital direct reading of very low resistances in the
alphanumerical display, with up to 4½ digits. Very fast
and accurate measurements.
:
RS-232 @ 4800 bps. Suitable for data collection in an
external serial printer, computer or data-logger.
:
IP54 with closed lid.