FIREBOY eco – Operating instructions V00
.
Error message Technical cause
Red LED is
flashing
(1x per sec),
green LED is on.
Red LED is on,
green LED is on.
Red LED
is flashing
quickly
(5x per sec),
green LED is on.
Red LED
is flashing slowly
(every 3 sec),
green LED is on.
Red LED is on,
green LED is off.
14
The flame monitor
does not detect
a flame.
The valve is closed.
Prior to ignition
process, the contact
pin on the burner head
is not detected.
Burner chamber
temperature
measured exceeds
the critical value.
Failure of
flame monitor.
Technical error.
Possible cause and correction
No gas supply is connected.
Connect the gas supply
Open the gas connection
Flame was blown out several
times.
Correctly set the gas/air
mixture
Prevent draft
Electrode of the flame monitor is
broken or dirty.
Visually inspect electrode
Clean electrode with alcohol
Burner head not added or not
added correctly.
Add burner head
Contact pin on burner head
is contaminated.
Clean the contact pin
Continuous operation overheated
the device.
Wait until device has
cooled off
Ambient temperature exceeds
the permitted value.
Reflection of the flame on the
device due to unauthorized use.
Electrode of the flame monitor
is kinked or dirty.
Visually inspect, if necessary
clean
Call service technician