Product specifications
Model:
Displayed temperature
range:
Operating ambient
temperature range:
Storage temperature:
Display resolution:
Accuracy for displayed
temperature range:
Clinical bias:
Limits of agreement:
Clinical repeatability:
Measuring Site:
Reference Site:
Operating and storage
relative humidity:
Battery life:
Service life:
Warranty:
If device is not used or stored within specified temperature and
humidity ranges the technical accuracy cannot be ensured.
This infrared thermometer meets requirements established
in ASTM Standard E 1965-98 (for the thermometer system
[thermometer with lens filter]). Full responsibility for the
conformance of the product to the standard is assumed by
Kaz USA, Inc.
ASTM laboratory accuracy requirements for the thermometer
only (not including a lens filter) in the display range of 96.8 °F
to 102.2 °F (36 °C to 39 °C) for infrared thermometers is ± 0.4 °F
(± 0.2 °C), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E 667-86
and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).
IRT 3030
93.2 °F – 108 °F (34 °C – 42.2 °C)
50 °F – 104 °F (10 °C – 40 °C)
-25°C - 55°C (-13°F - 131°F)
0.1°F or 0.1°C
±0.2°C in range 35-42°C (95-107.6°F)
±0.3°C outside that temperature range
-0.198
(-0.11
°F
± 0.45 °F ( 0.25 °C)
± 0.18 °F (0.10 °C)
Ear
Oral (estimated oral temperature
is displayed)
≤ 95% non-condensing
2 years /1000 measurements
5 years
3 years
English
°C)
13