values will be automatically stored.
3.4 Set Unit of Measure (Level 2)
TDR
Vp %Vc
VP=67%
Main 25Ω
Feet
Contrast
Shutdown: 3 minutes
Note: Vp is the correction factor for your cable needed to keep the
distance cursor accurate. 67% = 0.67. Experiment to get the best
value.
3.5 Set Cable Impedance (Z)
TDR
Vp %Vc
VP=67%
Main 25Ω
Feet
Contrast
Shutdown: 3 minutes
3.6 Set Velocity of Propagation (Vp)
Velocity of propagation may be set as % or speed in microseconds
(ms). The unit of measure, the speed is displayed in (feet or
meters), will be determined by the setting selected in section 3.4.
setting.