14. Technical specifications
Test currents
Accuracy of test current
Resistance ranges
Resolution
Output voltage
Measurement principle
Basic accuracyof
resistance measurement
Voltage measurement
Accuracy of voltage
measurement
Advanced features
Serial data output
Environmental protection
Safety class
Power supply
Built-in battery charger
: 1 mA - 10 mA - 100 mA - 1 A – 5 A - 10 A.
Each current may be continuously adjustable from
0 to 100%.
: ± 3%.
: 0 - 200 mΩ
@ 10 A
0 - 1000 mΩ @ 5 A
0 - 200 mΩ
@ 1 A
0 - 2000 mΩ @ 100 mA
0 - 20 Ω
@ 10 mA
0 - 2000 Ω
@ 1 mA
: 0.1 µΩ @ 10 A
: Up to 10 Vd.c. @ 1 A (open circuit).
: Four-terminal, Kelvin-type.
: ±(0.10% of reading + 3 ULSD*).
*
Units of the Least Significant Digit.
: 0 V - 5 V.
: ± 5%.
: Digital direct reading of very low resistances in the
alphanumerical display, with up to 4½ digits. Very
fast and accurate measurements.
: USB.
: IP54 with closed lid.
: Meets the requirements of IEC 61010-1
: Rechargeable battery 12 V – 6000 mAh or 220-
240 V~ mains supply.
: 220-240 V~ mains supply.
22