Gendex GXS-700
TM
Compatibility with Radiographic Generators
GXS-700 is generally compatible with any dental X-ray unit and generator capable of
supplying the required range of exposure times and doses.
Follow the instructions of the X-ray generator to set the desired dose range.
The radiation of X-ray tubes is controlled by the settings of:
Some controls allow for the modification of all of the above, some have fixed settings for
current and voltage.
Distance of X-ray Source to Sensor
There is a correlation between the distance of the cone to the sensor and the dose received by
the Gendex GXS-700 sensor. The radiation that reaches the sensor decreases with the square
of the distance. If you double this distance, you receive only 1/4 of the radiation.
Compatibility with Software
The GXS-700 requires the installation and use of GxPicture 3.0.1 or later software in
combination either with:
Important:
7-4
•
Exposure time (msec) or pulses
•
Voltage (kV or kVp)
•
Current (mA)
•
VixWin Platinum Software 2.0 or later, DEXIS 9.0.2 software or
later, or Dentrix Image 5.1 CU1 software or later.
or
•
Application software which has Dental IO Diagnostic Imaging
functionality; provided the software is released by the
manufacturer for usage with GxPicture software.
If the GXS-700 System is used with imaging software other than
VixWin Platinum, then the Intended Use of that software needs to
include dental intra-oral x-ray.
032-0279-EN Rev 1