Redox (O.R.P.) Calibration Menu
You can choose the automatic or manual mode.
Automatic calibration:
-
When the display shows the value of the buffer solution, insert the probe into the bottle, press
will show the countdown of the 60 seconds required to complete the calibration. If the quality of the alignment is
less than 50%, an error appears on the display; press
automatically exit the calibration); if the quality is higher than 50%, the value appears on the display; press
and the procedure is completed.
-
Manual calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press
will show the value 465mV flashing, insert the probe into my solution, press
solution, then press
EM00136304
to confirm and start the calibration procedure as described previously.
DosingPump.ir
to exit the calibration (after 4 seconds the pump will
rev. 1.0 EN
and the display
and the display
to display the value of my
11