Calibration:
The meter is calibrated at the factory (preset). Before use however check the right
setting with your system.
- For precise calibration, it is essential that the unit is free of air during the measuring process.
- Let the pump run for around 1 minute first.
- If possible, use an outlet pipe with a drip-stop valve.
- Use a calibrated measuring vessel with the largest possible capacity.
- Follow the steps in the description exactly.
1.
TOTAL
Press
and
The display will show
2.
Now put a specific quantity e.g. 5 or 10 litres into a calibrated measuring vessel.
For e.g. 10 litres the display will show
N.B.: The display „P 9999" must not be exceeded.
3.
Press RESET
The display will show
4.
Press TOTAL to change the flashing number from 0 to 9
RESET
Press
to confirm the set figure and to change to the next figure
Set the display to the amount just dispensed e.g. 10 litres.
The display will show
RESET
Press
to confirm the set figure
The electronics will now calculate the calibration factor
The display will first show
5.
RESET
Press
to confirm the calibration factor and to return to the normal mode
The display will show
Maintenance and care:
Fitting an in-line filter is recommended. If the display is faint change the batteries in good time.
Protect the meter from the effects of heat.
Warranty:
If the operating instructions are followed and the equipment used in accordance with the directions, we
accept liability within the bounds of the statutory warranty obligations (see our Delivery Conditions).
When making warranty claims always attach the invoice from your dealer, giving the date of purchase.
8
RESET
for min.10 seconds
P
L 0,00
(last character flashing)
L 10.00
- - - - -
and then e.g.
0.00
P 1500
00666