Press the Start Pushbutton . After 5 seconds, the display should
indicate a stable value, close to the nominal value of the standard
resistance.
If this were not the case, the accuracy of the instrument would be out
of its specification and it would be necessary to adjust by a qualified
laboratory.
If you want to calibrate the earth meter in other points of the present
range or of other ranges, adequate standard resistances must be
used. In order to avoid errors by the test leads resistances and their
connections, the 4-wire connection is used (fig 06), with the earth
meter in its measurement function of ρ (4 pole). According to the
scale, the highest acceptable error is:
Range
0 – 20.00 kΩ
0 – 2000 Ω
0 – 200 Ω
0 – 20 Ω
When using a test resistance higher than 100 Ω, the leads and contacts
resistances are not significant, thus figure 07 configuration can be used,
with the Function Selector Switch  in the R position (3 pole).
Maximum error allowed
±(2% of the measured value + 0.2 kΩ)
±(2% of the measured value + 20 Ω)
±(2% of the measured value + 2 Ω)
±(2% of the measured value + 0.2 Ω)