Gráficas de control de calidad para radiografía
general
Quality Control for
General radiography applications
Imager Type: __________
Film Type: ____________
Step 1: Print QC Test images on five consecutive days. Record the optical densities
measurements in the tables below. After five days, average the values to determine
the reference levels for each of the parameters.
Month
Day
Initials
Low Density
Mid Density
High Density
Average of 5 Values = calculated reference Hi g h D e nsi ty l e v el
Step 2: Copy the calculated reference levels to Charts 2A/B ('Daily Density Control Chart')
Figura 10: Gráfica 1: Determinación de niveles operativos
Drystar 5500, Drystar 5503 | Gráficas de control de calidad | 137
Determination of Operating Levels
Serial #: _________________
Emulsion #: ______________
Densitometer: ________________ (default selection)
Day 1
Day 2
Average of 5 Values = calculated reference L ow D e nsi ty l e v el
Average of 5 Values = calculated reference "Mid Density" level
Date ____________________
Input Tray: _______________
Day 3
Day 4
2901J ES 20220908 1526
Chart 1
Day 5