CRYSTAL analytical balance
The format of the euro protocol is:
D0 D1 D2 D3 D4 D5 D6 D7 D8 CR
1
2
3
4
5
D0 ... D8
Measure
CR
Carriage Return
LF
Line Feed
Is activated by the "PRINT" pushbutton. The output string uses 10 characters for the measure including sign
and decimal point, the value is right justified and filled with spaces from the left. The format of the output is:
D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B
1
2
3
4
5
Do to D9
Measure
B
Blank
U
Unit
CR
Carriage Return
LF
Line Feed
Note :
Data will be printed if and only if the wheight on pan is stable.
11.3. REMOTE CONTROL COMMANDS
The balance responds to the following commands:
Command
Function
Send data on display
B
C
Perform calibration
F
Stop continuous transmission of data on display
I
Start continuous transmission of data on display
T
Tare command
Every remote control command must end with <CR>.
Line-feed characters are considered dont'care symbols
INSRUCTION MANUAL - 60.50.630 9th issue - December 2014
EURO protocol
6
7
8
9
10
PRINTER protocol
6
7
8
9
10 11 12 13 14 15
LF
11
with sign & decimal point
ASCII code <CR>
ASCII code <LF>
U
U
U
CR LF
with sign & decimal point, right justified
Space between measure and unit
Measuring unit
ASCII code <CR>
ASCII code <LF>
ENGLISH
16
Page 31 of 45