Troubleshooting
Ensure the device inlet is not contaminated with sample liquid.
Otherwise clean it with a damp cloth, let it dry and try again.
Ensure there is no sample liquid on the windings or the rim of
the sample vials.
Ensure the device, vials, pipettes and your sample liquid are all
at ambient temperature.
Ensure you prepare the sample vial with exactly 2 mL of sample
liquid.
Perform a new calibration.
Try to clean the sensor (Settings → Advanced → Clean sensor).
Contact the Alivion technical support if the problem persists
Contamination warning does not go away
Make sure the device is operated in a room with clean air.
Ventilate the room. If necessary, remove any potential
contamination sources (e.g., open sample liquid, used sample
vials and pipettes).
Remove the O-ring from the device. Clean it and the sample vial
inlet with a damp cloth. Dry both and put the sealing O-ring
back in.
Try to clean the sensor (Settings → Advanced → Clean sensor).
Try using it repeatedly if necessary.
Contact the Alivion technical support if the problem persists.