Product Markings
Refer to Figure E1 and key below:
1. Certificate holder's name and address.
2. CE Mark and Notified Body Number (####).
3. UKCA Mark and Approved Body Number (####).
4. Pressure range. Example: 20 bar g (g: gauge; a:
absolute; sg: sealed gauge).
5. Date of manufacture (month/year).
6. Country of assembly: Made in UK/China.
7. Serial number.
8. ATEX certificate number.
9. UKEX certificate number.
10. European ATEX directive markings.
11. Hazardous area markings.
12. IECEx certificate number.
13. Warning text: "WARNING: Replace battery in safe area
only".
Installation
WARNING
Do not open the DPI104-IS when
an explosive atmosphere is present - this can
cause an explosion.
Do not use tools on the DPI104-IS that might
cause incendive sparks - this can cause an
explosion.
Do not connect an energized electrical circuit
in a hazardous area while explosive
atmospheres are present - this can cause an
explosion. Isolate the power supply to the
equipment first.
Batteries can cause incendive sparks. Install
the battery in DPI104-IS in a safe area only.
Use only Panasonic Industrial 6LR61AD.
CAUTION
Do not use force to turn the
pressure connector or the bezel farther than
the end stops. Using force can damage the
DPI104 / DPI104-IS.
Do not use the body of the DPI104 / DPI104-IS
to tighten the pressure connection, this can
cause damage. Use the flat faces on the
pressure connector to hold the body and
tighten the pressure union.
Batteries
CAUTION
Remove the battery from the unit
when placed in storage for an extended period
of time.
If the battery has leaked return the unit to an
approved Druck Service Centre.
For DPI104-IS use only 1 x 9V, IEC 6LR61,
ANSI/NEDA 1604 manganese alkaline primary cell from
the following list:
Panasonic Industrial 6LR61AD.
Replacing the Batteries
This symbol indicates that the battery is low.
The non intrinsically safe (Blue, DPI104) unit can operate
from a 9V, IEC 6LR61, ANSI/NEDA 1604 alkaline battery.
Copyright 2018 Baker Hughes Company.
2 | DPI104 / DPI104-IS Safety Instructions–English
The intrinsically safe (Yellow, DPI104-IS) unit shall only use
the batteries specified in section "Batteries" on page 2.
Fit as shown in Figure A1 using the following procedure:
1. If applicable, set the power to off and isolate the
external power supply.
2. Remove the display bezel (steps  and ).
3. Make sure that the o-ring (item 8) and the related
surfaces are serviceable. Use only original parts
supplied by the manufacturer.
4. Remove the battery clamp/cover (step , item 7).
5. Disconnect the battery connector (step , item 9) and
discard the used battery (step , item 10).
6. Use an applicable recycling facility.
7. Attach the battery connector (item 9) to the new
battery.
8. Install the new battery (item 10) and re-attach the
battery clamp/cover (item 7).
9. Push the display bezel back into position until it is fully
engaged.
Specific Conditions of Use
When the power supply is through the 8-pin connector, use
only a Type A or a Type B cable as specified in
IEC 60079-14.
Declaration Requirements – EU Directive 2014/34/EU
This equipment is designed and manufactured to meet the
essential health and safety requirements not covered by
EU Type Examination Certificate ITS07ATEX25517X when
installed as detailed in this guide.
Declaration Requirements – UK SI 2016/1107 (as
amended by SI 2019/696)
This equipment is designed and manufactured to meet the
essential health and safety requirements not covered by
UK-Type Examination Certificate ITS21UKEX0076X when
installed as detailed above.
Pressure Safety
This equipment meets the requirements of the European
Pressure Equipment Directive 2014/68/EU and
UL 61010-1 for pressure safety.
Return Goods/Material Procedure
If the unit requires calibration or is unserviceable, return it
to the nearest Druck Service Centre listed at:
https://druck.com/service.
Contact the Service Department to obtain a Return
Goods/Material Authorization (RGA or RMA). Provide the
following information for a RGA or RMA:
Product (e.g. DPI104 / DPI104-IS)
Serial number.
Details of defect/work to be undertaken.
Calibration traceability requirements.
Operating conditions.
Quick Start
Before Using the Instrument
Read and understand the "Safety" section.
Check the DPI104 / DPI104-IS for signs of damage. Do
not use a damaged DPI104 / DPI104-IS.
Power On Sequence
Refer to Figure B1 and the procedure below.
1. Press the
button.