Descargar Imprimir esta página

College Park CAPITAL Manual Del Usuario página 3

Rodilla hidráulica

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 12
ENVIRONMENTAL CONDITIONS
Operating Temperature: -1 - 38° C (30 - 100° F)
USING THE COLLEGE PARK CAPITAL KNEE
The College Park Capital Knee has three functional modes. To change the mode, press down on the mode switch lock, then
rotate the switch to the left, center, or right position. When lined up, the lock will click into place.
LOCK MODE
When the switch is moved to the left,
the knee provides high resistance to
flexion. It will continue to extend until
full extension is reached.
Caution: To prevent sudden collapse, users should use caution when applying weight to the knee during free
swing mode..
NORMAL WALKING
MODE
At the center position, the knee functions
according to the hydraulic settings of
swing flexion, swing extension, and stance
flexion resistance.
FREE SWING MODE
When the switch is moved to the right,
stance flexion is bypassed. This allows free
movement in both flexion and extension.
EN

Publicidad

loading