O
PERATION
o Place the tubes into the tube holders. Be sure to follow the rules for balanced loads as listed on page 8.
o The front panel LED is illuminated for the currently selected cycles. The selected cycle determines the run time
and speed. To select another cycle, press the UNLOCK button in succession until the desired cycle is selected.
Note: cycle selection is only available with the lid open.
o Close the lid and turn the lid knob clockwise to its complete stop position.
o Pushing the START button on the control panel will start the spin cycle.
o When the cycle is completed, the rotor will slow to a complete stop and the lid light will flash.
o The lid will unlock for 60 seconds allowing entry into the rotor chamber. To unlock after more than 60 seconds
have elapsed, press the UNLOCK button. The lid will unlock for another 15 seconds.
o With the lid unlocked, turn the lid knob counterclockwise and open the lid. The lid light will turn off.
o You may now safely remove the samples.
S
ETTINGS
S
S
TANDARD
ETTINGS
HORIZON 6
HORIZON 12
HORIZON 24
Blue Tube Holders
HORIZON 6
75 & 100 mm tubes
FA
Orange Tube
Holders 125 mm
tubes G-Force
Drucker Diagnostics – Customer Service: +1-814-692-7661 –
(1) Chemistry
RPM
Time (Mins)
G-Force
RPM
Time (Mins)
G-Force
RPM
Time (Mins)
G-Force
RPM
Time (Mins)
G-Force
(2) Coag (PPP)
3,600
10
1,800 xg
3,500
10
1,800 xg
3,300
10
1,800 xg
3,900
10
1,600 xg
1,850 xg
CustomerService@DruckerDiagnostics.com
3,300
15
1,500 xg
3,200
15
1,500 xg
3,000
15
1,500 xg
3,800
15
1,500 xg
N/A
(3) Urine
1,900
5
500 xg
1,800
5
500 xg
1,700
5
500 xg
2,200
5
500 xg
600 xg
| Page 7