The Dräger Aerotest 5000 automatically calculates the pressure
dewpoint from the entered water concentration and the system
pressure. The system compares this with the ambient temperature to
determine whether the test has been passed or not.
This test result is automatically recorded in the memory of the Dräger
Aerotest 5000. The test results are recorded according to the date
and time at which the corresponding test was completed. Up to 10
tests can be recorded. To view the test results, select the "Review
tests" menu from the main menu. Previous tests can be viewed or
deleted in this menu.
To disconnect the Dräger Aerotest 5000 from the air supply to be
tested, make sure that the inlet coupling remains connected to the
Dräger Aerotest 5000 and only disconnect it from the end of the air
supply hose. The adapters can then be removed from the Dräger
Aerotest 5000.
3.3.3
Testing an air supply of 200 bar or 300 bar (high pressure
systems)
The F 3002 pressure reducer with an inlet coupling for an input
pressure of 300 bar max. must be used for all high pressure tests.
When checking compressed air cylinders, the cylinder to be tested
must be completely filled.
If the compressed air cylinder is not completely filled, the measured
water concentration may be incorrect.
Select the required test program:
High-pressure cylinder ≤ 200 bar,
High-pressure cylinder > 200 bar,
High-pressure charging compressor.
Then select the method for the oil measurement and follow the menu
for the further procedure.
Dräger Aerotest 5000
Make sure that the flow control valve is closed, then connect the
supplied hose to the pressure reducer and the air supply connection of
the Dräger Aerotest 5000. Next connect the F 3002 pressure reducer
to the cylinder to be tested, or the compressor. Use the connection
hose provided.
Open the cylinder valve of compressed air cylinders and check that the
compressed air cylinder is completely filled. The pressure indicated in
the display corresponds to the set cylinder pressure and must not
exceed 10 bar.
NOTICE
i
i
If the Dräger Aerotest 5000 is over-pressured an
overpressure warning is displayed and the unit will no longer
be able to perform tests. The Dräger Aerotest 5000 must be
sent to Dräger to be tested and readjusted.
Open the flow control value momentarily and smell the air, if a foul odor
can be detected at the sintered air outlet (6) stop the test and press the
"Fail" key. Seriously contaminated air could result in damage of the
Dräger Aerotest 5000. If no foul odor can be detected, press the
"Pass" key.
The Dräger Aerotest 5000 will commence a purge sequence of
approx. 5 minutes to remove contaminations from previous
measurements.
Once the purge has finished, break off the tips of the selected Dräger-
Tubes and insert them into the corresponding tube sockets (2-5). The
arrow on the tube points away from the device. Manually tighten the
tube socket on the device.
The Dräger Oil-Impactor is inserted into the socket for the oil tube.
Use
19