Descargar Imprimir esta página

Thuasne BP 3MS1-4N Manual De Instrucciones página 17

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
8. Using a mains adapter
You can operate this device using the Thuasne mains adapter (DC
6V, 600mA).
Only use the Thuasne mains adapter available as an orig-
inal accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket 8 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
9. Error messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat
the measurement.*
«ERR 2»
Error signal
During the measurement, error signals
were detected by the cuff, caused for
BK
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
Thuasne BP 3MS1-4N
Error
Description Potential cause and remedy
«ERR 3»
No pressure
An adequate pressure cannot be
in the cuff
generated in the cuff. A leak may have
BL
occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces-
sary. Repeat the measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate
result
and no result can therefore be
displayed. Read through the checklist
for performing reliable measurements
and then repeat the measurement.*
«ERR 6» MAM Mode
There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for performing
reliable measurements and then repeat
the measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high
pressure too
(over 299mmHg) OR the pulse is too
high
high (over 199 beats per minute). Relax
for 5 minutes and repeat the measure-
ment.*
«LO»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measure-
ment.*
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
10.Safety, care, accuracy test and disposal
m
Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
EN
15

Publicidad

loading

Productos relacionados para Thuasne BP 3MS1-4N