English
EN
GUIDELINES AND DECLARATION OF THE MANUFACTURER
1) This equipment needs to be installed and put into service in accordance with the information provided in the
ACCOMPANYING DOCUMENTS;
This product needs special precautions regarding EMC and needs to be installed and put into service according
to the EMC information provided, and this unit can be affected by portable and mobile RF communications
equipment.
2) Caution: Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may
result in incorrect operation of the unit.
3) Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
4) Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or
stacked use is necessary, this machine should be observed to verify normal operation in the configuration in
which it will be used.
Guidance and manufacture's declaration – electromagnetic emission
The Device is intended for use in the electromagnetic environment specified below. The customer of the
user of the Device should assure that it is used in such an environment.
Emission test
Compliance
RF emissions
Group 1
CISPR 11
RF emission
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
Not applicable
IEC 61000-3-3
Electromagnetic environment – guidance
The Device use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The Device is suitable for use in all establishments,
other than domestic and those directly connected
to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
32
Guidance and manufacture's declaration – electromagnetic immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user
of Device should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
±6 kV contact
(ESD)
±15 kV air
IEC 61000-4-2
Electrical fast transient/
±2 kV for power
burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to earth
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
Voltage dips, short
for 5 cycles
interruptions and voltage
variations on power supply
70% UT
input lines
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
English
Electromagnetic
Compliance level
environment - guidance
Floors should be wood, concrete
or ceramic tile. If floor are
±6 kV contact
covered with synthetic material,
±15 kV air
the relative humidity should be at
least 30%.
Mains power quality should be
Not applicable
that of a typical commercial or
hospital environment.
Mains power quality should be
Not applicable
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Device requires continued
Not applicable
operation during power mains
interruptions, it is recommended
that the Devicebe powered from
an uninterruptible power supply
or a battery.
33