Miele MF 01 RU Instrucciones De Montaje página 12

Módulo 1 de dosificación líquida del lado descontaminadocontaminado 1-6

Publicidad

Produktgruppe 512
12 von 46
Note
If the dispensing display is activated, the operator can activate and deactivate dispensing. If the
dispensing display is not activated, dispensing activation and deactivation is controlled by the
programme in operation only.
To quit the service mode, switch off the machine.
Connect the dispenser pumps in accordance with the wiring diagram and calibrate them as follows:
If the pump capacity (flow rate) is known, enter the figure directly in the supervisor level as follow (e.g. for
dispenser pump 1).
Supervisor level
Main menu
OK button
Back button (to go back one level)
If the pump capacity (flow rate) is unknown, calibrate the dispenser pump in the supervisor level as follows
(e.g. for dispenser pump 1).
Supervisor level
Main menu
Place the dispenser pump suction tube in a container with a level mark.
Fill the container to the level mark.
Press button 0/1 (dispenser pump operates for 60 s).
Use a measuring vessel to refill the container to the level mark.
Note the dispensed quantity via the measuring vessel and enter this figure in the display (V:.....ml).
Activate the disk symbol to save the setting.
Register the empty signal as follows (e.g. for dispenser pump 1).
Supervisor level
Main menu
button
Back button (to go back one level)
Check dispensing in a programme and carry out adjustments if necessary.
Carry out a functional check.
Carry out a data backup via the service mode.
Service level
Backup
disk symbol to overwrite EW data.
To quit the service mode, switch off the machine.
Carry out final work:
Check the machine for leaks.
Refit the lid parts and cover plates.
Carry out appropriate electrical safety checks.
Reconnect the machine to the electric mains supply.
en
09.06.2010
Diese Unterlagen dürfen ohne unsere Genehmigung weder vervielfältigt noch Dritten zugänglich gemacht werden. Eigentumsrechte vorbehalten.
Settings
Dispensing capacity
Settings
Dispenser calibration
Settings
Dispenser empty signal
Activate disk symbol to save setting.
Backup data to EW (selection module)
Umbau- und Montageanweisung
Dispenser pump 1: No
Activate disk symbol to save setting.
Dispenser pump 1:
Dispenser pump 1: No
Overwrite existing data?
M.-Nr. 07503821
.....ml/min
Yes
OK
Activate