Connecting And Installing Cartridges; Connection Options; Connect The Cartridge Using The Screw Connectors; Connecting Cartridges With Quick Release Couplings - Miele professional PG 8595 Instrucciones De Manejo

Aqua-purificator
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 45
Connect the car‐
tridge using the
screw connec‐
tors.
Connecting car‐
tridges with quick
release couplings
en - Connecting and installing cartridges

Connection options

Open the door by pulling it upwards and disengage it from the hin‐
ges to remove it.
Unscrew the safety caps from the cartridge connectors.
Connect the hoses according to the colour coding on the cartridge
connectors,
"Inlet (Eingang)" (red) to "In" ,
"H2O pur" (green) to "Out" .
The cartridges and hoses can be fitted with quick release couplings
(optional accessory) for rapid changing and reconnection.
When installing for the first time, screw the male part of the quick re‐
lease couplings to the threaded cartridge connectors (in-out).
Check for leaks before using for the first time.
Connect the hoses with the relevant colour coding to the cartridge
couplings and press down until they click firmly into place.
To release a quick release coupling pull the ring on the end of the
hose upwards and pull the hose off.