en - Connecting and installing cartridges
Venting the car‐
tridges
Inserting a car‐
tridge
40
Connecting cartridges to the mains water supply and
the washer-disinfector
Guide the hoses through the openings provided on the back of the
Aqua Purificator.
Ensure that the hoses are not kinked when doing this.
Connect the "H2O pur" (green) hose to the solenoid valve for AD
water or to the AD water inlet hose for the washer-disinfector.
Connect the "Inlet (Eingang)" (red) hose to the on-site cold water
mains supply stopcock.
Important
Water pressure (flow pressure) must be between 250 - 1000 kPa
pressure (2.5 - 10 bar).
The water temperature must not exceed 20 °C.
Open the valve by turning it to the left.
Open the cold water stopcock until water escapes from the valve.
until water escapes from the vent valve.
Then close the valve.
The demineralisation cartridge is now ready for use.
Place the connected cartridges into the back of the Aqua Purifica‐
tor.
The front area can be used to accommodate a spare cartridge.