Warning And Safety Instructions - Miele professional A 105/1 Instrucciones De Manejo

Cesto superior del inyector
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 48
en - Warning and Safety instructions
38
Read the operating instructions carefully before using this upper
basket.
Keep these operating instructions in a safe place.
 Please also read the operating instructions for your washer-
disinfector and pay particular attention to the Warning and Safety
instructions.
The upper basket is only approved for use in the applications
stated in these operating instructions. Components such as nozzles
may only be replaced with Miele accessories or genuine original
spare parts.
New mobile units and baskets must be washed in the washer-dis-
infector without a wash load prior to first use.
Inspect all mobile units, baskets, modules and inserts as described
in the "Maintenance" section of the operating instructions for your
cleaning machine.
For your own safety, wear gloves and, if necessary, a surgical
mask and protective goggles when sorting the instruments and un-
loading the machine. To avoid recontamination, wear clean gloves
when unloading the machine.
Load from back to front and unload from front to back.
To clean and disinfect lumened instruments thoroughly from the in-
side, they must be attached to the relevant holders or nozzles on the
injector manifolds. The connectors must not come loose during a
programme sequence.
At the end of every programme sequence, check that the connectors
have not come loose.
Any hollow items that have become disconnected from their
injector nozzles, adapters or holders during processing must be
re-processed.
The last rinse, as a minimum, must be carried out with DI water.

Publicidad

Tabla de contenido
loading

Tabla de contenido