Installation

Where necessary, the nozzles supplied can be replaced with different
fittings, e.g. holders for dental transmission instruments, irrigation
sleeves, etc. Further examples are given on the following pages.
 The wash water used for reprocessing narrow lumen instru-
ments needs to be filtered. The injector manifolds can each be fit-
ted with an A 800 tubular filter for this purpose. The tubular filter is
supplied with its own operating instructions.
Cleaning the A 800 tubular filter
The tubular filter must be cleaned at least once a week.
We recommend cleaning the tubular filter after every 10 programme
sequences.
Follow the cleaning instructions in the operating instructions for the
tubular filter.
A 803 holder for transmission instruments
A 803 holders must be retro-fitted for reprocessing dental transmis-
sion instruments. The holders are suitable for almost all standard
transmission instruments.
Remove a nozzle from the injector manifold and replace it with an
A 803 holder.
When the A 803 holders are used, the wash water must be filtered.
The injector manifold must be fitted with an A 800 tubular filter.
The tubular filter is supplied with its own operating instructions.
Follow in particular the instructions on how to perform regular
cleaning on the tubular filter.
en - Areas of application
43