"Output Connector Not For Current lnterrupting".
For continued protection against risk of fire, replace
only with same type and ratings of fuse.
To reduce the risk of fire and electric shock, install in
a temperature- and humidity-controlled indoor area
relatively free of conductive contaminants.
Power Supply Cord: Use UL Listed detachable power
supply cord-No. 18 AWG, 2 conductors, flexible cord,
rated 10 A, VW - 1, 105 °C, minimum 1.8 m, maximum
3 m long. Provided with a molded–on, non–polarized
attachment plug with a 15 A, 125 V (NEMA1-15P)
configuration and a molded-on connector which mates
with the power inlet.

1.5 Operating instructions

1. Make sure the battery charger output voltage is the
same as the output voltage of the connecting battery.
2. Connect the battery charger to the charging socket of
the mobility device.
3. Connect the power cord. The LED indicates red and
green flash when AC power is on.
4. Start charging. Refer to 1.3 LED display, page 4 .

1.6 Troubleshooting

If the POWER LED (red) is off:
– Check that the charging cable is correctly connected.
– If the LED still does not light, the battery charger
may be faulty. Contact your provider.
1654662-A
If the CHARGING LED is off:
– Check that the charging cable is correctly connected.
– If the battery is fully charged, the charger will switch
to trickle charge mode and the CHARGING LED light
goes out.
– If the charging process was not started (orange LED),
the battery could be faulty. Contact your provider.
If the green CHARGING LED keeps flashing, it cannot
turn to indicate charging:
– Check if the battery is connected successfully.
– Check if the output connection is short or open.
– If the battery connection is ok, the battery charger
may be faulty.
If the POWER LED (red) keeps flashing:
– Check if the battery connection is reversed.
– Check if the output connection is short or open.
– Check if the environment temperature is too low (<
0 °C)
– If the POWER LED keeps flashing, the battery charger
may be faulty.
If the CHARGING LED does not change from orange
to green:
– The battery cannot be charged correctly. It may be
faulty. Stop charging and contact your provider.
If the CHARGING LED changes from orange to green
immediately:
– The battery is either fully charged or faulty. Contact
your provider.