Table 4 Warning list for conductivity sensors (continued)
Warning
WRONG LINEAR TC
WRONG TC TABLE
WRNG USER CONC TABLE
WRNG BLT-IN TEMP TABLE
WRNG BLT-IN CONC TABLE

Replacement parts and accessories

Refer to the replacement parts and accessories section of the controller documentation for controller
parts and accessories.
Note: Product and article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts and accessories
Description
Kit of 3 protection caps for instrument front panel connectors
Kit of 2 adapters D6/8 to DN4/6
Kit of 2 black protection caps for sample input and output connectors
Power supply connector
Tool for disconnecting in/out sampling tubes
DN8 semi-rigid PTFE tubing (per meter)
DN8 semi-rigid PE tubing (per meter)
Cable for the 4-20 mA output (5 meters)
Cable for the 4-20 mA output (10 meters)
Cable for the 4-20 mA output (20 meters)
¾'' NPT flow chamber in PP with fittings
Pt100 temperature simulator (0.1°C accuracy)
Cable for connecting the Pt100 temperature simulator
Yearly re-calibration in our factory
Description
The user-defined linear temperature
compensation is out of range
The user-defined temperature
compensation table is out of range
The concentration measurement is
outside of the range of the user
table
The measured temperature is
outside of the range of the built-in
temperature compensation table
The concentration measurement is
outside of the range of the built-in
concentration table
Resolution
The value must be between 0 and
4%/°C; 0 to 200 °C.
The temperature is above or below
the temperature range defined by
the table.
Make sure that the user table is set
for the correct measurement range.
Make sure that the temperature
compensation is configured
correctly.
Make sure that the concentration
measurement is configured for the
correct chemical and range.
Item no.
09126=A=8010
09126=A=8020
09126=A=8030
350=500=004
578=507=602
590=060=080
151400,22387
08319=A=0005
08319=A=0010
08319=A=0020
09126=A=0100
037=000=001
09125=A=8020
09526=A=1000