15. manufacturer's data regarding emc
electromagnetic emissions:
the device is suitable for use in the electromagnetic environment indicated below. the
user must ensure that the device is used in such an environment.
test
result / complied with
hF-emission
Group 1
hF-emission
class B
harmonic
yes
emissions
Flicker
yes
emission
Voltage
yes
fluctuations
electromagnetic stability (immunity):
the device is suitable for use in the electromagnetic environment indicated below. the
user must ensure that the device is used in such an environment.
test
test step
electrostatic
+/-8 kV air +/-6
discharge
kV contact
Voltage peaks
+/-2 kV
mains +/-1
kV input
ri-vital
®
ManuFacturer's Data reGarDInG eMc
74
comment
the device must radiate electromagnetic energy
in order to fulfil its function. the functioning of
surrounding devices can be affected by this.
Iec 61000-4-2
the device is suitable for use in all facilities.
Iec 61000-4-4
Iec 61000-3-2
Iec 61000-3-3
Iec 61000-3-3
classification
comment
+/-8 kV air +/-6
the floor should be made of a non-conductive
kV contact
material.
the relative humidity should be at least 30%.
Iec 61000-4-2
+/-2 kV
the mains power should correspond to the power
mains +/-1 kV
in a typical business environment.
kV input
Output
Output
current surge
+/- 1kV difference
+/-1kV difference
+/- 2kV common
+/- 2kV common
mode Iec 61000-4-5
mode
Voltage drops
<5% un(> 95%
drop in ut)
see test step
test
test step
classification
Voltage drops
40% un (60%
see test step
drpo in ut) with
five full changes.
Voltage drops
70% un (30%
see test step
drop in ut) with
25 full changes
Voltage drops
<5% un (60%
see test step
drop in ut) in
five seconds
Power frequency 3a/m
3a/m
ri-vital
®
ManuFacturer's Data reGarDInG eMc
Iec 61000-4-4
the mains power should correspond to the power in
a typical hospital or business environment.
the mains power should correspond to the power
in a typical hospital environment. If uninterrupted
operation must absolutely be guaranteed, then an
uninterrupted power supply must be used.
comment
the mains power should correspond to the power
in a typical hospital environment. If uninterrupted
operation must absolutely be guaranteed, then an
uninterrupted power supply must be used.
the mains power should correspond to the power
in a typical hospital environment. If uninterrupted
operation must absolutely be guaranteed, then an
uninterrupted power supply must be used.
the mains power should correspond to the power
in a typical hospital environment. If uninterrupted
operation must absolutely be guaranteed, then
an uninterrupted power supply must be used. Iec
61000-4-11
In the event of operating problems, the device must
be placed at a greater distance to the source of the
magnetic fields. Iec 61000-4-8
note: un is the mains voltage for the test device
75