8.3
Checking the System Joint's Free Movement
Check if the system joint moves freely. If the system joint runs with lateral play, mount the next thicker sliding
washer. If it does not move freely (it is jammed), mount the next thinner sliding washer.
8.4
Securing the Screws
The screw is secured after the orthosis has been produced and tried on and before it is handed over to the
patient.
1 Secure the screw for the joint's upper part (fig. 9) with the torque corresponding to the system width and
LOCTITE® 243 medium strength.
2 Let the adhesive harden (final strength after approx. 24 hours).
Screw for Joint's Upper Part
S1 (screw 1)
The screw of the joint's upper part is not secured with the necessary torque at delivery. You can also
find information on the torque on the cover disc of the system joint.
9.
Mounting to the System Anchor
The system anchor of the NEURO CLASSIC H
the system joint according to our recommended production technique (fig. 10). You will
find more detailed information in the Instructions for Use for Orthotists or Qualified/
Trained Experts System Anchors for Waterproof System Joints. You will find informa-
tion on the production techniques in the section "Online Tutorials" on our website
www.fior-gentz.com.
20
O system knee joint has to be adhered to
2
System Width
20mm
4Nm