In case of malfunctions please pay attention to the following points:
Trouble-
Problem: No or low nebulization
Shooting
Solution: Replace the batteries, clean the electrodes
slightly moistened, smooth cloth and clean the appliance according
to chapter „4.1 Cleaning maintenance". Check, whether the
nebulizer unit is correctly mounted on the device base and if there is
sufficient contact between inhalant and mesh (hold device straight).
Problem: No nebulization and power indicator
or lights up only shortly.
Solution: Check, if the batteries have been inserted correctly and if
these provide enough power. If in doubt, exchange the batteries.
Check, whether the nebulizer unit is correctly mounted on the device
base and if there is sufficient contact between inhalant and mesh
(hold device straight).
Problem: The device switches itself off during, although there is still
inhalant in the tank.
Solution: The mesh
between nebulizer unit and device base. Clean the electrodes
with a slightly moistened, smooth cloth and check, if the nebulizer
unit has been mounted onto the device base correctly and if there is
sufficient contact between inhalant and mesh (hold device straight).
Exchange the nebulizer unit.
Problem: Inhalant leaks off the device.
Solution: The silicone seal could be defective. Check, if the
nebulizer unit is correctly mounted on the device resp. exchange the
nebulizer unit.
If above points do not help to solve the problem, contact the
customer service.
4.3
This inhalation device meets the requirements of the EU standard
Directives
93/42/EWG (medical device directive) and of the standard EN
/ norms
13544-1: 2007 Respiratory therapy equipment - Part1: Nebulising
systems and their components.
Electromagnetic compatibility:
The device complies with the EN 60601-1-2 standard for
electromagnetic compatibility. You can find details on this
measurement data on the seperate enclosure.